Vaisala HMP240 User Manual

Page 33

Advertising
background image

HMP240 series

M210300en

User's Guide

31

5.3.5.4

Selecting the output units

Select Unit in the main menu:

Use the arrow keys to select the right alternative and press ENT.

metric

non-metric

RH

%RH

%RH

T

°C

°F

Td

°C

°F

dT

°

C

°

F

a

g/m

3

gr/ft

3

x

g/kg

gr/lb

Tw

°C

°F

5.3.5.5

Selecting the calculation mode

Disconnect the security lock jumper!

Select More and then again More in the second menu:

Select Frost and then the desired alternative with the arrow keys;

FROST ON (default) for frostpoint and FROST OFF for dewpoint
calculation at dewpoint temperatures below 0

°

C.

5.3.6 Output modes

The output modes only affect output through the serial interface: the transmit-
ter accepts all display and LED commands irrespective of which serial output
mode it is in. The HMP240 transmitter has three serial output modes: RUN,
STOP and POLL.

In the RUN mode the transmitter outputs measurements automatically through
the serial interface to a PC or a peripheral. The only command that can be
given through the serial interface is S (stop), which ends the RUN mode.

In the STOP mode serial commands are given to the transmitter. Measure-
ments are then output only by entering command SEND.

The POLL mode is used when more than one transmitter is connected to the
same serial bus; a single transmitter can be addressed and communicated with.