Vaisala HMP240 User Manual

Page 126

Advertising
background image

HMP240
Appendix 7: Re-gaining

M210300en

124

Use arrow keys to change the status of the automatic re-gaining and press
ENT.

Acknowledge the current setting by pressing ENT. To change the interval
press CL. Select the new time interval digit by digit with the arrow keys and
acknowledge each digit with ENT key. With CL key you can correct the entry
by deleting the digits one by one. When you have entered the whole time
interval, press ENT for the second time.

Other re-gaining parameters are acknowledged or changed the same way.
When all the parameters have been changed or acknowledged, the menu
Regen. settings/Regeneration is displayed again.

4.2.

Starting re-gaining manually

To start re-gaining manually select Regeneration and press ENT. The
following is displayed

When the heating is completed or it has been aborted, the settling period
begins and the transmitter returns to the display mode outputting the measured
or calculated values. However, during the settling time or until the
temperature difference is small enough, the display is locked to the values
measured before re-gaining was started. A new re-gaining cannot be started
during the settling period.