Vaisala HMP240 User Manual

Page 57

Advertising
background image

HMP240 series

M210300en

User's Guide

55

Enter the measured upper end current/voltage on channel 2.

6.3.3 With LED commands

If both the analogue outputs and humidity/temperature channels are calibrated,
the analogue outputs should be calibrated first. This applies only when the
calibrations are done with LED commands!

connect an ammeter/voltmeter to the analogue outputs (connector X2)

Disconnect the security lock jumper!

Give command ¡¡ll.

the LED on the left flashes; set the low end of channel 1 with the ar-
row keys and press ENT

the second LED from the left flashes; set the high end of channel 1
with the arrow keys and press ENT

the LED on the left flashes; set the low end of channel 2 with the ar-
row keys and press ENT

the second LED from the left flashes; set the high end of channel 2
with the arrow keys and press ENT

The analogue outputs are calibrated to ensure the correct scaling: for example,
when the output is scaled to 4...20 mA, the low end of the scale is 4 mA and
high end 20 mA exactly. However, when 0... 20 mA output is used, the output
can not be adjusted to exactly 0 mA, but to 50 µA. When 0...1 V, 0...5 V or
0...10 V output is in use, the output is adjusted to 50 mV. The following table
summarizes the correct output values.

Summary of the correct output values:

Output scale:
0...20 mA

4...20 mA

0...1 V

0...5 V

0...10 V

low end:

50 µA

4 mA

50 mV

50 mV

50 mV

high end:

20 mA

20 mA

1 V

5 V

10 V