Appendix 1: Serial commands

75

CTA

Temperature calibration of the optional T sensor head

Disconnect the security lock jumper!

CTA <cr>

With this command the optional temperature sensor head can be calibrated
against an accurate reference, such as a Pt 100 simulator. A two-point
calibration is performed as follows:

>CTA <cr>
Ta : 0.80

Ref1 ?

0.0 <cr>

Press any key when ready ...
Ta : 56.20

Ref2 ?

55.0 <cr>

In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:

>CTA <cr>
Ta : 0.80

Ref1 ?

0.0 <cr>

Press any key when ready ...
Ta : 75.50

Ref2 ?

<cr>

If the stabilization of the sensor to the temperature of the calibrator or the ref-
erence needs to be monitored, the measurement output can be repeated by giv-
ing command c<cr> at Ref1 and Ref2:

>CTA <cr>
Ta : 0.80

Ref1 ?

c <cr>

Ta : 0.40

Ref1 ?

0.00 <cr>

Press any key when ready ...
Ta : 56.20

Ref2 ?

55.0 <cr>

ACAL

Calibrating the analogue outputs

Disconnect the security lock jumper!

ACAL <cr>

Calibrates the outputs selected on channels 1 and 2. The output is measured
and the measured values (mA or V) entered as calibration coefficients.

For example, calibrating the outputs when 0...10 V signal has been selected on
both channels (set with AMODE command)

>ACAL <cr>
Ch1

U1

(V ) ?

0.123 <cr>

Ch1

U2

(V ) ?

9.98 <cr>

Ch2

U1

(V ) ?

0.120 <cr>

Ch2

U2

(V ) ?

9.98 <cr>