HMP240
Appendix 7: Re-gaining

M210300en

126

1.2.3 CRI

Changing re-gaining parameters

CRI <cr>

With command CRI the values of the re-gaining parameters can be changed.

Example: changing the re-gaining interval to 480 min and the settling time

after sensor heating to 360 s

>cri
Interval min : 720 ? 480
Duration s : 120 ?
Temp diff : 0.50 ?
Settl time s : 240 ? 360
k % : 95.00 ?
>

NOTE

To activate new parameter settings immediately,
command RESET must be given. Otherwise the
transmitter starts the re-gaining with new parameters
only when the old re-gaining interval is completed.

1.2.4 FST

Displaying re-gaining status

FST nnn <cr>

nnn = ON or OFF

With FST ON the status of the re-gaining is output at the beginning of the
output line.