8.

TECHNICAL DATA

Sensor

RH & T composite sensor
HUMICAP

®

KC(part no.

18258HM)

T sensor in the humidity sensor head

Pt 100 IEC 751 class 1/3 B

Voltage proof, AC: DC supply (+ or -) to housing 250 VAC, 1 minute
(300 k

and 15 nF parallel)

8.1

Dewpoint temperature

typical range of dewpoint temperature -40...+100 °C

Response time (90 %) at +20 °C in

15 s

still air (with sintered filter)

0

1

2

3

4

5

6

0

20

40

60

80

100

Dewpoint difference (°C)

Difference between ambient temperature and dewpoint temperature

Accuracy in dewpoint

measurement (°C)

-20

0

20 40

60

-40

Dewpoint temperature (°C)

100

80

Figure 8.1

Accuracy in dewpoint measurement as a function of

depression and dewpoint (NOTE: dewpoint < 0