Vaisala HMP240 User Manual

Page 21

Advertising
background image

HMP240 series

M210300en

User's Guide

19

CH2 CURRENT OUTPUT

CH1 0 ... 1 VOLTAGE OUTPUT

OPENED COVER OF THE HMP 243

CH1

CH2

C

H

2

C

H

1

X15

CURRENT OUTPUTS
0 ... 20 / 4 ... 20 mA

VOLTAGE OUTPUTS
0 ... 5 V / 0 ... 10 V

VOLTAGE OUTPUTS
0 ... 1 V

CH1

CH2

C

H

2

C

H

1

CH1

CH2

C

H

2

C

H

1

CH1

CH2

C

H

2

C

H

1

Figure 4.2

Selecting the analogue outputs with jumpers

The software has to be informed which outputs are in use. This is done either
through the serial interface or the menus on a local display. The serial
command is AMODE and the display/keypad command "Mode ð Analog
outputs ð Mode" (see Chapter 5). For the scaling of the outputs, see serial
command ASCL and the display command "Mode ð Analog outputs ð
Scale".

All the jumpers are used only with the 0...1 V outputs. When other outputs are
in use, the spare jumpers are kept in connector X55.

Advertising