Appendix 1: Serial commands

M210300en

82

Activating the date output

>fdate on
Form. date : ON
>r
1995-03-10 RH= 21.1 %RH T= 26.0 'C
1995-03-10 RH= 21.3 %RH T= 26.0 'C
1995-03-10 RH= 23.1 %RH T= 26.0 'C
1995-03-10 RH= 22.2 %RH T= 26.0 'C
1995-03-10 RH= 20.6 %RH T= 26.0 'C
...

Inactivating the date output

>fdate off
Form. date : OFF
>r
RH= 20.2 %RH T= 26.0 'C
RH= 19.9 %RH T= 26.0 'C
RH= 19.8 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
...

SERI

Serial bus settings

SERI b p d s x <cr>

b

=

bauds (300, 600, 1200, 2400, 4800, 9600)

p

=

parity (n = none, e = even, o = odd)

d

=

data bits (7 or 8)

s

=

stop bits (1 or 2)

x

=

duplex (H = half, F = full)

Giving the command on its own outputs the current settings:

>SERI <cr>
4800 E 7 1 FDX

The settings can be changed one parameter at a time or all parameters at once:

>SERI O H <cr>

changing parity and duplex

4800 O 7 1 HDX

>SERI 600 N 8 1 F <cr>

changing all parameters

600 N 8 1 FDX

NOTE

The serial bus settings become effective only after reset.