Vaisala HMP240 User Manual

Page 83

Advertising
background image

HMP240

M210300en

Appendix 1: Serial commands

81

?\UUU.U\\t\TTT.T\\t\DDD.D\\t\AAA.A\\t\XXX.X\\t\WWW.W\\
r\n <cr>

The output is then:

47.4<tab> 22.4 <tab> 10.6 <tab> 9.4 <tab> 8.0 <tab>
15.4 <cr><lf>

FTIME

Adding time to output

FTIME xxx <cr>

xxx

=

ON or OFF

When FTIME is activated, the current time is output at the beginning of the
output line. The time is set with command TIME. After RESET or power on
the current time is 00:00:00.

Activating the time output

>ftime on
Form. time : ON

>intv 5 s

setting the output interval

Output intrv. : 5 s
>r
09:31:13 RH= 19.4 %RH T= 26.0 'C
09:31:18 RH= 19.4 %RH T= 26.0 'C
09:31:23 RH= 19.8 %RH T= 26.0 'C
09:31:28 RH= 19.6 %RH T= 26.0 'C
09:31:33 RH= 19.5 %RH T= 26.0 'C
09:31:38 RH= 19.5 %RH T= 26.0 'C
...

Inactivating the time output

>ftime off
Form. time : OFF
>r
RH= 19.4 %RH T= 26.1 'C
RH= 19.8 %RH T= 26.1 'C
RH= 20.6 %RH T= 26.1 'C
RH= 20.5 %RH T= 26.1 'C
RH= 19.9 %RH T= 26.1 'C
RH= 19.6 %RH T= 26.1 'C
...

FDATE

Adding date to output

FDATE xxx <cr>

xxx

=

ON or OFF

When FDATE is activated, the current date is output at the beginning of the
output line. The time is set with command DATE. After RESET or power on
the current date is 1991-01-01.

Advertising