Vaisala HMP240 User Manual

Page 41

Advertising
background image

HMP240 series

M210300en

User's Guide

39

The time can be set in seconds within the range of 0 - 1024 (0 = no averaging
time).

For example:

>FILT
Filter (S):

0 ? 1024

>

>FILT 100 <cr>
Filter (S):

100

>

5.4.2.8

Setting the transmitter address

ADDR aa <cr>

aa =

address (0...99)

Example: transmitter is given address 99

>ADDR <cr>
Address : 2 ? 99 <cr>

5.4.2.9

Setting the calculation mode

Disconnect the security lock jumper!

FROST ON/OFF <cr>

This command is used to select whether the transmitter calculates the
frostpoint (default) or the dewpoint reading at dewpoint temperatures below
0

°

C. Select FROST ON for frostpoint and FROST OFF for dewpoint

calculations. For example:

>Frost <cr>
Frost :

ON

>Frost off
Frost : OFF
>Frost on
Frost : on
>

5.4.2.10 Resetting the transmitter

RESET <cr>