Vaisala HMP240 User Manual

Page 55

Advertising
background image

HMP240 series

M210300en

User's Guide

53

6.2.2.2

With display / keypad commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

Select Cali in the main menu and then T (Ta for the
additional T sensor head); select two-point calibration T
(or Ta) 2 point cal. Change the first point reading with the
arrow keys and press ENT.

•

Change the temperature and check the transmitter again
against the reference.

•

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

•

If necessary, change the second point reading with the ar-
row keys and press ENT.

6.2.2.3

With LED commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

Connect an ammeter/voltmeter to the analogue outputs
(connector X2). Give command ¡¡l¡ (or ¡l¡¡ for
the T sensor head). At the first calibration point the LED
on the left flashes; adjust the first point (offset) with the
arrow switches to the same reading with the reference and
press ENT switch.

•

Change the temperature and check the transmitter again
against the reference.

•

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

•

If necessary, adjust with the arrow switches to the correct
value and press ENT. At the second calibration point the
second LED from the left flashes.