6.2.2.2

With display / keypad commands

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

Check the transmitter against the reference.

Select Cali in the main menu and then T (Ta for the
additional T sensor head); select two-point calibration T
(or Ta) 2 point cal. Change the first point reading with the
arrow keys and press ENT.

Change the temperature and check the transmitter again
against the reference.

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

If necessary, change the second point reading with the ar-
row keys and press ENT.

6.2.2.3

With LED commands

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

Check the transmitter against the reference.

Connect an ammeter/voltmeter to the analogue outputs
(connector X2). Give command ¡¡l¡ (or ¡l¡¡ for
the T sensor head). At the first calibration point the LED
on the left flashes; adjust the first point (offset) with the
arrow switches to the same reading with the reference and
press ENT switch.

Change the temperature and check the transmitter again
against the reference.

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

If necessary, adjust with the arrow switches to the correct
value and press ENT. At the second calibration point the
second LED from the left flashes.