5.4.3.2

OPEN & CLOSE

OPEN nn <cr>

nn

=

address of the transmitter (0...99)

CLOSE <cr>

In STOP mode: command OPEN has no effect, CLOSE sets the transmitter in

POLL mode

In POLL mode: command OPEN sets the transmitter temporarily in STOP

mode, command CLOSE returns the instrument to POLL
mode

Example: relative humidity calibration is performed at transmitter 2 which is

in POLL mode

>OPEN 2 <cr>

opens the line to transmitter 2

>CRH <cr>

calibration started

...

>CLOSE <cr>

line closed