Vaisala HMP240 User Manual

Page 12

Advertising
background image

HMP240-series
User's Guide M210300en

10

a p l u g g e d h o l e f o r r e f e r e n c e
m e a s u r e m e n t s

T - s e n s o r h e a d

( o p t i o n a l )

h u m i d i t y s e n s o r h e a d

d u c t w a l l

s e a l i n g ( s i l i c o n e )

f l a n g e

P T F E s l e e v e

s u p p o r t i n g b a r

Figure 3.2

Installing the sensor heads of the HMP243 in a channel
with the help of flanges and supporting bars

The sensor head can also be installed vertically.

m

h

h

t o b e s e a l e d

( s i l i c o n e )

h u m i d i t y s e n s o r

h e a d

T - s e n s o r h e a d

s t a i n

o r a

P T F E - s l e e v e

a b u s h i n g ( c a b l e - g l a n d )

( s e a l : V i t o n )

( o p t i o n a l )

Figure 3.3

Vertical installation of the HMP243 sensor heads

when a bushing is used, its
size is selected according to
the diameter of the sensor
head; the diameter of the
cable is increased by using
e.g. silicon tape at the bushing

stainless steel
cable tie or a
similar fastener

Advertising