The output scales can also be given directly with the ASCL command.

ASCL aaa.a bbb.b ccc.c ddd.d <cr>

aaa.a

=

lower limit of channel 1

bbb.b

=

upper limit of channel 1

ccc.c

=

lower limit of channel 2

ddd.d

=

upper limit of channel 2

For example, when relative humidity is scaled to 0...100 %RH on channel 1
and temperature to -40...+100

°

C on channel 2:

>ASCL 0 100 -40 100 <cr>
Ch1

(RH) lo

0.000 %RH

Ch1

(RH) hi

100.000 %RH

Ch2

(T )

lo

-40.000 'C

Ch2

(T )

hi

100.000 'C