Vaisala HMP240 User Manual

Page 129

Advertising
background image

HMP243

M210300en

Appendix 7: Re-gaining

127

Example: activating status display

>fst on
Form. status : ON
>r

N 0 RH= 20.8 %RH T= 25.6 'C

N = re-gaining off

N 0 RH= 20.7 %RH T= 25.6 'C
N 0 RH= 20.7 %RH T= 25.6 'C
N 0 RH= 20.8 %RH T= 25.6 'C

H 163 RH= 20.8 %RH T= 25.6 'C

H = heating on, 163

H 163 RH= 20.8 %RH T= 25.6 'C

describes the heating

H 163 RH= 20.8 %RH T= 25.6 'C

power

H 163 RH= 20.8 %RH T= 25.6 'C

S 163 RH= 20.8 %RH T= 25.6 'C

S = settling period on

S 163 RH= 20.8 %RH T= 25.6 'C
S 163 RH= 20.8 %RH T= 25.6 'C
S 163 RH= 20.8 %RH T= 25.6 'C
N 163 RH= 20.8 %RH T= 25.6 'C
N 163 RH= 15.0 %RH T= 31.5 'C
N 163 RH= 16.7 %RH T= 30.0 'C
N 163 RH= 17.4 %RH T= 29.1 'C

H 160 RH= 17.7 %RH T= 28.6 'C

H = new heating (new

H 160 RH= 17.7 %RH T= 28.6 'C

power)

deactivating status display

>fst off
Form. status : OFF
>r
RH= 20.1 %RH T= 26.4 'C
RH= 20.2 %RH T= 26.4 'C
RH= 20.1 %RH T= 26.4 'C
RH= 20.2 %RH T= 26.4 'C

1.2.5 REGEN

Starting re-gaining manually

REGEN <cr>

Re-gaining can also be started manually with command REGEN. The current
re-gaining parameters (duration and settling times stored with command CRI)
are used.

Example: starting re-gaining

>regen
Regeneration started, press any key to abort
>

The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to values measured before re-gaining until the
settling time is over.