Vaisala HMP240 User Manual

Page 78

Advertising
background image

HMP240
Appendix 1: Serial commands

M210300en

76

L

Outputting linear correction coefficients

L <cr>

With the help of command L the user can check how the transmitter has been
adjusted after it has been calibrated at the factory.

>L <cr>
RH

offset

: 0.000

RH

gain

: 1.000

Ts

offset

: 0.000

Ts

gain

: 1.000

Ta

offset

: 0.000

Ta

gain

: 1.000

where: Ts = temperature of the humidity sensor head

Ta = temperature of the temperature sensor head

LI

Entering linear correction coefficients

Disconnect the security lock jumper!

LI <cr>

The LI command is one way of calibrating the transmitter.

>LI <cr>
RH

offset

: 0.000 ? 0.1 <cr>

RH

gain

: 1.000 ? <cr>

Ts

offset

: 0.000 ? <cr>

Ts

gain

: 1.000 ?

Ta

offset

: 0.000 ? -0.05

Ta

gain

: 1.000 ?

where: Ts = temperature of the humidity sensor head

Ta = temperature of the temperature sensor head

The factory settings are offset 0 and gain 1. The transmitter can be returned to
its factory calibration by giving these values .

NOTE

The temperature unit in offset correction is always de-
grees Centigrade, even if the transmitter is using non-
metric units (Fahrenheit) in its measurement output.