6.2.2

Two point temperature calibration

6.2.2.1

With serial commands

•

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the reference.

•

For the humidity sensor head: give command CT <cr>,
enter the first point value and press <cr>:

>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready

•

If you want to see how the sensor stabilizes to the refer-
ence temperature, enter c <cr> instead of the first refer-
ence:

T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...

•

For the additional temperature sensor head: give command
CTA <cr>, enter the first point value and press <cr>:

>CTA<cr>
Ta: -20.07 Ref1 ? -20.08

Press any key when ready...

•

Change the temperature and check the transmitter again
against the reference.

•

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

•

Press any key, enter the second point value and press <cr>.

T (or Ta) : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by
entering c <cr> instead of the reference value.