Vaisala HMP240 User Manual

Page 54

Advertising
background image

HMP240
User's Guide

M210300en

52

6.2.2

Two point temperature calibration

6.2.2.1

With serial commands

Leave the reference instrument and the transmitter for at
least 4 hours in the same space so that their temperatures
have time to equalize. Remove the filter cap prior to
calibration.

Disconnect the security lock jumper!

Check the transmitter against the reference.

For the humidity sensor head: give command CT <cr>,
enter the first point value and press <cr>:

>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready

If you want to see how the sensor stabilizes to the refer-
ence temperature, enter c <cr> instead of the first refer-
ence:

T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...

For the additional temperature sensor head: give command
CTA <cr>, enter the first point value and press <cr>:

>CTA<cr>
Ta: -20.07 Ref1 ? -20.08

Press any key when ready...

Change the temperature and check the transmitter again
against the reference.

Check that the reading corresponds to the reading of the
reference instrument. If not, adjust the second point.

Press any key, enter the second point value and press <cr>.

T (or Ta) : xx.x Ref2 ? yy.y <cr>

The stabilization of the sensor can be monitored by
entering c <cr> instead of the reference value.