Vaisala HMP240 User Manual

Page 85

Advertising
background image

HMP240

M210300en

Appendix 1: Serial commands

83

The processor does not allow the following combinations:

no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2

even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1

When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.

UNIT

Selecting the output units

UNIT x <cr>

x

=

m(etric units) n(on-metric units)

metric units

non-metric
units

RH

%RH

%RH

T

°

C

°

F

Td

°

C

°

F

dT

°

C

°

F

a

g/m

3

gr/ft

3

x

g/kg

gr/lb

Tw

°

C

°

F

For example, the command for setting the non-metric units is:

>UNIT N <cr>
Output units : non metric

When the command is given with no parameters, the transmitter outputs the
currently valid setting.

ADDR

Setting the transmitter address

ADDR aa <cr>

aa

=

address (0...99)

The address is used when more than one transmitter is connected to one serial
bus. The ADDR command makes it possible to communicate with one
transmitter at a time in POLL state.

For example, transmitter is given address 99

>ADDR <cr>
Address : 2 ? 99 <cr>