Insert the sensor head into the measurement hole of the
LiCl chamber.

•

Wait at least 10 minutes for the reading to stabilize.

•

Give command CRH <cr>, enter the first point value and
press <cr>.

>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...

•

If you want to see how the sensor stabilizes to the humid-
ity in the calibrator, enter c <cr>:

RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

•

Insert the sensor head into the measurement hole of the
NaCl chamber.

•

Wait at least 20 minutes for the reading to stabilize.

•

Press any key, enter the second point value and press
<cr>.

RH : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by
entering c<cr>.

6.1.2.2

With display / keypad commands

•

Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the transmitter.

Disconnect the security lock jumper!

•

Insert the sensor head into the measurement hole of the
LiCl chamber.

•

Wait at least 20 minutes for the reading to stabilize.

•

Select Cali in the main menu and then RH cal; select Not
changed and then two-point calibration RH 2 point cal.
Change the first point reading with the arrow keys to cor-
respond to the reference humidity and press ENT; pressing
an arrow once changes the reading by 0.05 %RH.