>SERI O <cr>

changing parity only

4800 O 7 1 HDX

>SERI 600 N 8 1 F <cr>

changing all parameters

600 N 8 1 FDX

The processor does not allow the following combinations:

no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP240 programme will change the number of stop bits to 2

even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1

NOTE

The serial bus settings become effective only after reset.

When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.

5.4.2.6

Selecting the output units

UNIT x <cr>

x =

m(etric units)
n(on-metric units)

metric

non-metric

RH

%RH

%RH

T

°C

°F

Td

°C

°F

dT

°C

°F

a

g/m

3

gr/ft

3

x

g/kg

gr/lb

Tw

°C

°F

5.4.2.7

Setting the averaging time

Disconnect the security lock jumper!

FILT nnnn <cr>

nnn

=

averaging time (0 - 1024 seconds)

This command is used to set and inspect the averaging time during which the
individual measurement samples are integrated to get an averaged reading.