Serial commands – Vaisala HMP240 User Manual

Page 127

Advertising
background image

HMP243

M210300en

Appendix 7: Re-gaining

125

5. SERIAL COMMANDS

1.2.1 REG

Re-gaining settings

REG <cr>

REG nnn <cr>

nnn = ON or OFF

Command REG is used to check the re-gaining parameter settings and to
activate or inactivate the automatic re-gaining.

Examples: checking the current settings

>reg

REG OFF

automatic re-gaining is inactive

Interval min : 720
Duration s : 120
Temp diff : 0.50
Settl time s : 240
k % : 95.00
>

activating the automatic re-gaining

>reg on
REG ON
>

deactivating the automatic re-gaining

>reg off
REG OFF
>

1.2.2 CR

Re-gaining parameters

CR <cr>

Command CR can be used to check the values of re-gaining parameters.

Example:

>cr
Interval min : 720
Duration s : 120
Temp diff : 0.50
Settl time s : 240
k % : 95.00
>