Appendix 5: Calculation formulas

117

=

Θ

+

Θ

=

3

1

4

ln

ln

i

i

i

ws

b

b

P

where:

bi

= coefficients

b-1 = -0.58002206 * 104

b0

= 0.13914993 * 101

b1

= -0.48640239 * 10-1

b2

= 0.41764768 * 10-4

b3

= -0.14452093 * 10-7

b4

= 6.5459673

The parameters A, m, and T

n

depend on temperature according to the follow-

ing table:

t

A

m

T

n

0 ... 50 °C

6.1078

7.5000

237.3

50 ... 100 °C

5.9987

7.3313

229.1

100 ... 150 °C

5.8493

7.2756

225.0

150 ... 180 °C

6.2301

7.3033

230.0

When T

d

< 0 and the transmitter is set to dewpoint calculation mode (FROST

OFF), the parameters are as follows:

A

= 6.119866

m

= 7.926104

Tn

= 250.4138

When T

d

< 0 and the transmitter is set to frostpoint calculation mode (FROST

ON), the parameters are as follows:

A

= 6.1134

m

= 9.7911

Tn

= 273.47