Appendix 3: Digital current loop module

107

>send 22
RH= 24.4 %RH T= 29.1 'C

•

Disconnect the terminal.

•

Remount the digital current loop module.

•

Close the cover.

•

Repeat this setting procedure with each transmitter

•

When all transmitters on the network have been configured, switch
them off.

CHECKING THE SERIAL PORT NETWORK
OPERATION

Measurement readings are normally asked with the transmitter in POLL mode;
the command must include the address of the transmitter. If the settings of the
transmitter need to be changed, the transmitter is switched to STOP mode with
command OPEN; commands can then be given without address. When the line
to the transmitter is closed, it returns to POLL mode.

STOP mode

Open the line to the transmitter:

open 22<cr>
HMP 22 line opened for operator commands

Transmitter no. 22 is now temporarily set to STOP mode; it accepts com-
mands without address until CLOSE command is given. Individual settings
can now be easily modified. Do not open more than one line at a time.

Use command ? to find out the settings of the active transmitter:

? <cr>
HMP 240 / 1.02
CPU serial nr : 0
Keyboard type : 0
Address

: 7

Output units

: metric

Baud P D S

: 4800 E 7 1 FDX

Serial mode

: STOP

Output intrv. : 0 min
Mtim

: 32

Pressure : 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH )

lo

0.000 %RH

Ch1 ( RH )

hi

100.000 %RH

Ch2 ( T )

lo

-20.000 'C