Checking the serial port network operation, Stop mode – Vaisala HMP240 User Manual

Page 109

Advertising
background image

HMP240

M210300en

Appendix 3: Digital current loop module

107

>send 22
RH= 24.4 %RH T= 29.1 'C

Disconnect the terminal.

Remount the digital current loop module.

Close the cover.

Repeat this setting procedure with each transmitter

When all transmitters on the network have been configured, switch
them off.

CHECKING THE SERIAL PORT NETWORK
OPERATION

Measurement readings are normally asked with the transmitter in POLL mode;
the command must include the address of the transmitter. If the settings of the
transmitter need to be changed, the transmitter is switched to STOP mode with
command OPEN; commands can then be given without address. When the line
to the transmitter is closed, it returns to POLL mode.

STOP mode

Open the line to the transmitter:

open 22<cr>
HMP 22 line opened for operator commands

Transmitter no. 22 is now temporarily set to STOP mode; it accepts com-
mands without address until CLOSE command is given. Individual settings
can now be easily modified. Do not open more than one line at a time.

Use command ? to find out the settings of the active transmitter:

? <cr>
HMP 240 / 1.02
CPU serial nr : 0
Keyboard type : 0
Address

: 7

Output units

: metric

Baud P D S

: 4800 E 7 1 FDX

Serial mode

: STOP

Output intrv. : 0 min
Mtim

: 32

Pressure : 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH )

lo

0.000 %RH

Ch1 ( RH )

hi

100.000 %RH

Ch2 ( T )

lo

-20.000 'C