WIKA CPC6000 User Manual

Page 23

Advertising
background image

Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

23

Control Time

Fast Mode: 10 seconds to stable flag for 10% FS step
pressure change into a 50cc volume. Larger volumes
can lengthen this time. Controlling to low absolute
pressures will lengthen this time.
Slow Mode: 15 seconds to stable flag for 10% FS step
pressure change into a 50cc volume. Larger volumes
can lengthen this time. Controlling to pressures less
than 0.5 psia will lengthen this time.

Supply Consumption

Fast Mode: Zero after setpoint is reached. All supply
gas is used for upscale pressure sample filling.
Slow Mode: Zero

Measure to Control Offset

Fast and Slow mode: <0.0005% Span

Overshoot

<1% of active range.

4.3 Control Specifications for the Solenoid Valve Regulator Module

External Pressure Requirements:

Source Requirements

10% over range of highest pressure transducer or 20
psi over highest pressure transducer for a pressure
channel, whichever is less.

Exhaust Requirements

A vacuum source is required for sub-atmospheric
pressure control.

Stability of Controlled Pressure

0.003% of span on active range. Typically better than
0.001% of span 10 seconds after displaying stable
flag.

Sensor Spans Available in each Regulator
Type

LP SVR: 0-1 psig to 0-50 psig (0-5 psia to 0-50 psia).
MP SVR: 0-10 psi to 0-150 psi.
HP SVR: 0-75 psi to 0-1500 psi.

Minimum Control Pressure

0.05% FS or .025 psi over exhaust pressure, which-
ever is greater.

Control Time

10 seconds to stable flag for 10% pressure step
change into 50cc volume. Larger volumes can
lengthen this time. Controlling to pressures less than
0.5 psia can also lengthen this time.

Pressure Control Rates

Slow, Medium, Fast, Max (default).

Supply Consumption

<2.5 scfh in steady-state control.

Overshoot

Low overshoot mode: <.005%
High speed mode: <1%