Mensor/WIKA Operating Instructions - CPC 6000

6.3.8 Step Down [

›] and Step Up [š] Keys

As shown in Figure 6.3A, the [

›] and [š] keys are located to the side (different for A and B Channels) of

the [Setpoint Value] key; one for Step Down [

›], and another for Step Up [š]. Pressing an arrow key will

change the setpoint by the step value until the control limits of the channel are reached.

6.3.9 Control Pressure [Setpoint Value]

The number displayed inside the [Setpoint Value] key is the setting for a controlled pressure output.
When the [Control] key at the bottom of the active channel screen is switched to [On] the regulator will
attempt to present that precise pressure to the Measure/Control port on the rear panel. The setpoint
number is changed either by using the [

š] and [›] keys, or by touching the [Setpoint Value] key to input

a new number.

If the displayed number is beyond the range of a selected turndown in range hold mode, the number
will change to a control value within the limits of the turndown. The setpoint will change in value auto-
matically, but it will not be restored automatically if the previous turndown is reselected.

[Standby] key: When this key is ON all internal solenoid valves are closed.

[Measure] key: When this key is ON the active sensor and the Measure/Control port are connected
and isolated from the rest of the internal pneumatics.

[Control] key: When this key is ON the unit will attempt to control to the setpoint displayed.

[Vent] key: When this key is ON the internal pneumatics are open to atmosphere.