Automated Pressure Calibrator
CPC 6000

50

Mensor/WIKA Operating Instructions - CPC 6000

Always send commands in one of the following formats:

1.

[Command] [Termination];

2.

[Command] [Separator] [Data] [Termination];

3.

Queries are special instructions in the form:
[Command?] [Termination] where the question
mark, “?”, immediately precedes the terminator.

When a valid query is received, the CPC 6000 will return {data} terminated by CR and LF.

Floating point data is returned in the current engineering units in exponential format.

7.4.4 Output Formats

Pressure readings are returned in exponential notation in a format according to the OUTFORM com-
mand as follows:

1. <sp> pressure value <cr><lf>
2. <sp> pressure, units number, mode <cr><lf>
3. <sp> pressure, pressure rate <cr><lf>
4. <sp> pressure, minimum peak, maximum peak<cr><lf>
5. <sp> pressure, active sensor (P or S), active turndown (1-4)<cr><lf>
6. <sp> pressure, control point, “stable” or “slewing”<cr><lf>
7. <sp> pressure, “no barometer” or baro reading <cr><lf>

The space <sp> at the beginning of the string is replaced with an “E” if the CPC 6000 has detected a
remote communication error.

For commands that take boolean data the following strings are acceptable:

0

1

False

True

No

Yes

Off

On