WIKA CPC6000 User Manual

Page 60

Advertising
background image

Automated Pressure Calibrator
CPC 6000

60

Mensor/WIKA Operating Instructions - CPC 6000

7.4.7 Remote Emulation

The Mensor PCS 400 is an earlier generation instrument similar to the CPC 6000. There is some
compatibility between the CPC 6000 and a PCS 400 in that the CPC 6000 will respond to many of the
remote instructions as if it were the older instrument. The PCS 400 commands will operate only on the
currently active control channel.

Table 7.4.7.1 is a list of the remote commands and queries which the CPC 6000 will recognize and
respond to.

A space between elements in a command indicates a required delimiter. Use a space where such a
delimiter is indicated. A full description of the syntax and use for each of these commands and queries
are presented in the PCS 400 manual.

i

Notice

All pressure values will be in the currently active pressure units
unless otherwise stated.

7.4.7.1 PCS 400 Emulated Commands and Queries

Table 7.4.7.1 - PCS 400 Emulated Commands and Queries

Ref

Command

Data

Function / Response

1

_pcs4 autorange <value>

0 or 1

1 turns autorange on, 0 turns autorange off.

2

_pcs4 autorange?

Returns 1 if in autorange, 0 if in range hold.

3

_pcs4 cal span <value>

Sets the span of the active transducer to <value>.

4

_pcs4 cal zero <value>

Sets the zero of the active transducer to <value>.

5

_pcs4 cal_disable_off

Enables zero or span calibrations if previously disabled.

6

_pcs4 cal_disable_on

Prevents zero or span calibrations.

7

_pcs4 ctrl <value>

Sets control value. Will take effect immediately if instru-
ment is in control mode.

8

_pcs4 ctrl?

Returns the current control point in current engineering
units.

9

_pcs4 ctrlmax <value>

Sets maximum control value.

10

_pcs4 ctrlmax?

Returns current maximum control pressure.

11

_pcs4 ctrlmin <value>

Sets minimum control value.

12

_pcs4 ctrlmin?

Returns current minimum control pressure.

13

_pcs4 default

Sets default values into instrument.

14

_pcs4 err?

Returns the error number and description.

15

_pcs4 func ctrl <value>
<unitno>

Instrument placed in control mode at <value> pressure
in <unitno> units.

16

_pcs4 func F1

Toggles between absolute and gauge modes if an inter-
nal barometric sensor is installed.

17

_pcs4 func meas

Instrument placed in measure mode.