Automated Pressure Calibrator
CPC 6000

64

Mensor/WIKA Operating Instructions - CPC 6000

:MODE OFF/GAS/LIQUIT

Sets the head correction mode.

:MEDIUM <n>

Sets the density of the medium.

SYSTem
:DATE <i,i,i>

Sets the system date (YY,MM,DD).

:TIME <i,i,i>

Sets the system time (hh,mm,ss).

:ERRor[:NEXT]?

Return: error#, “description”.

:KLOCk ON/OFF/1/0

Lock or unlock keyboard.

:PRESet Load known values.
:SAVe

Saves settings to non-volatile memory.

:VERSion?

Return of the SCPI-standard.

TEST
:ELECtronic?

Returns electronics status.

:RELay<n>?

Returns whether relay 1 to

3

is closed or not.

:RELay<n> ON/OFF

Turns relay 1 to

3

on or off.

UNIT
:NAME<n>?

Returns the name of the unit with index n.

:FACTOR<n>?

Returns the conversion factor of the unit with index n.

index unit
0 bar
1 mbar
2 Pa
3 psi
4 atm
5 kp/cm2
6 lbf/ft2
7 kPa
8 cmH2O(4°C)
9 inH2O(4°C)
10 inH2O(60°F)
11 ftH2O(4°C)
12 μmHg(0°C)
13 mmHg(0°C)
14 cmHg(4°C)
15 inHg(0°C)
16 inHg(60°F)
17 - -
18 user
19 user
20 user

OUTPut
:STATe ON/OFF/1/0

Turns control mode on or off.

:STATe? Returns control mode “0” for off, ”1” for on.
:MODE MEASure/CONTrol/VENT

Sets the operation mode.

:MODE?

Returns current operation mode MEASURE”,

CONTROL”or “VENT”.

:STABLE?

Returns 1 if in control mode and stable.