WIKA CPC6000 User Manual

Page 58

Advertising
background image

Automated Pressure Calibrator
CPC 6000

58

Mensor/WIKA Operating Instructions - CPC 6000

7.4.6 CPC 6000 Error Codes

Table 7.4.6 - CPC 6000 Error Codes

Code

Serial
Poll
Byte

Description

Error String Returned

E00

00h

reserved

unused code

E01

41h

GPIB listen/talk error

GPIB LISTEN/TALK ERROR

E02

42h

Syntax error

UNKNOWN COMMAND

E03

43h

Expected a valid _PCS4 command

EXPECTED A VALID _PCS4 COMMAND

E04

44h

Expected a valid FUNC command

EXPECTED A VALID FUNC COMMAND

E05

45h

Expected a valid CAL command

EXPECTED A VALID CAL COMMAND

E06

46h

Expected a valid TEST command

EXPECTED A VALID TEST COMMAND

E07

47h

Expected a pressure units selection

EXPECTED A PRESSURE UNITS SELEC-
TION OR INVALID TERMINATION STRING

E08

48h

Expected a pressure value

EXPECTED A PRESSURE VALUE

E09

49h

Expected a transducer selection

EXPECTED A XDUCER SELECTION

E10

4ah

Expected a seconds selection

EXPECTED A SECONDS SELECTION

E11

4bh

reserved

unused code

E12

4ch

reserved

unused code

E13

4dh

Invalid pressure units selection

INVALID PRESSURE UNITS SELECTION

E14

4eh

Invalid control pressure value selection INVALID CONTROL PRESSURE VALUE

SELECTION

E15

4fh

Invalid rate value selection

INVALID RATE VALUE SELECTION

E16

50h

Invalid a/d unit selection

INVALID A/D UNIT SELECTION

E17

51h

Invalid zero offset value selection

INVALID ZERO OFFSET VALUE SELECTION

E18

52h

Invalid Span offset value selection

INVALID SPAN OFFSET VALUE SELECTION

E19

53h

Invalid rate units selection

INVALID RATE UNITS SELECTION

E20

54h

Sensor overrange

SENSOR OVERRANGE

E21

55h

Sensor underrange

SENSOR UNDERRANGE

E22

56h

Sensor failure detected

SENSOR FAILURE DETECTED

E23

57h

Low source pressure

LOW SOURCE PRESSURE

E24

58h

Regulator failure detected

REGULATOR FAILURE DETECTED

E25

59h

reserved

unused code

E26

5ah

Internal leak detected

INTERNAL LEAK DETECTED

E27

5bh

Program error detected

PROGRAM ERROR DETECTED

E28

5ch

Memory error detected

MEMORY/COEFFICIENT ERROR DETECT-
ED

E29

5dh

External Leak detected

EXTERNAL LEAK DETECTED

E30

5eh

High exhaust pressure

VACUUM ERROR DETECTED

E31

5fh

Transducer error detected

XDUCER ERROR DETECTED

Advertising