Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

41

6.6.2 [Sensor] Setup

Touch the [Sensor] tab and Figure 6.6.2A will appear.

Figure 6.6.2A - Sensor Setup Screen

Filter: The Filter is an electronic filter to smooth out the pressure readings. Because of the differences
in resolution, more filtering may display a more stable reading for some pressure units. Select the best
filter for the current units. [Off], [Low], [Normal], [High].

Resolution: The Resolution [value] key allows the user to enter the number of significant digits that will
be displayed on the operate screen. See Figure 6.6.2B.

Figure 6.6.2B - Resolution Window

[Calibrate] Setup key: Details for the [Calibrate] key are given in

Section 10

, Calibration.

6.6.3 [Controller] Setup

The controller setup screen (Figure 6.7.1A for the pump regulator or Figure 6.8 for the solenoid valve
regulator) is used to set the control parameters for the selected pressure control channel.

The controller test screens are an interactive diagnostic display used for troubleshooting the overall
pneumatic system of the CPC 6000. Proper use of these features are described in

Section 9

, Mainte-

nance.