Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

67

usage “I1”

dummy function accepts the string and returns what was entered

usage “I2”

dummy function accepts the string and returns what was entered

usage “I3”

dummy function accepts the string and returns what was entered

usage “I4”

dummy function accepts the string and returns what was entered

usage “I5”

dummy function accepts the string and returns what was entered

usage “I6”

dummy function accepts the string and returns what was entered

case ‘J’:

/*Rate_Mode*/

usage “J0”

sets the rate mode to variable rate mode

usage “J1”

sets the rate mode to variable rate mode

usage “J2”

sets the rate mode to regular control mode (default) MAX

Note: All of the notation formats have a leading space!

case ‘N’:

/*Notation_Mode*/

usage “N0”

output format default responds with pressure/remote/range/scale/
display/error

usage “N1”

output format responds with pressure or setpoint depending on the
display mode setting

usage “N2”

output format responds with range/scale/display/control mode/IO/
function/error

usage “N3”

output format returns 0 for not stable or 1 for stable

usage “N4”

output format returns error/error(on/off)/rate/variable rate/units

N0

if display is set to 1 the pressure reading is the setpoint reading
example 200.00REMR1S3D1@01
if display is set to 0 the pressure reading is the pressure reading
example 199.98REMR1S3D0@01

N1

if display is set to 1 the pressure reading is the setpoint reading
example 200.00
if display is set to 0 the pressure reading is the pressure reading
example 199.98

N2

example R1S3D1C0I0F00@01

N3

example 1

N4

@01@1J0V+00009U BAR

case ‘P’:

/*Pressure_Setpoint_Mode*/

usage “PXXXX.
XXXX”

X = sets the pressure setpoint (ex. P13.5)

case ‘R’:

/*Range(xducer)*/

usage “R0”

no range

usage “R1”

range 1 if available (main range)

usage “R2”

range 2 if available (secondary range)

case ‘S’:

/*Scale*/

usage “S1”

sets the scale with one unit selection

usage “S2”

sets another scale with another unit selection

usage “S3”

sets the global scale enables all units to be selected