Automated Pressure Calibrator
CPC 6000

38

Mensor/WIKA Operating Instructions - CPC 6000

[Setpoint%] key: The [Setpoint%] key sets the control pressure at the entered percentage of the
range of the currently active transducer.

[Delay] key: The [Delay] key delays the execution of the program for the entered number of seconds.

[Wait] key: The [Wait] key delays the execution of the program until the instrument measures a stable
pressure, or the control pressure stabilizes within the control parameters or pauses until the operator
presses a key to continue the program execution.

[Start] key: The [Start] key causes execution of the program to begin at the first program line. This
is useful for running a program repeatedly until the stop key (the key with the black filled square) is
pressed by the operator.

6.4.2 Running a Program

To run a stored program, select the program mode from the channel setup screen (see “6.6.1 - [Chan-
nel] Setup”). The main operation screen looks like Figure 6.4.2.

Figure 6.4.2 - Main Operating Screen in Program Mode

At the bottom of the screen, the function tabs (Measure, Control and Vent) are replaced by information
and tabs pertaining to the selected program. The large key at the top of this section is the program
name. Any saved program may be selected directly from this screen by pressing this key and selecting
the program to run.

When a program has been selected, the line below the program name key shows the current program
line that is or will be executed. The rightmost box on this line shows the time at which the program has
been on the current program line.

The five keys below this control the operation of the program.

The [<<] key steps one command line back.

The [>>] key steps one command line forward.