94

Rsetpt?

<sp>{value}<cr><lf>

Returns the current rate setpoint of the active
channel in current units per second.

95

Sbaud

{9600, 19200, 38400, 57600}

Sets the serial baud rate.

96

Sbaud?

<sp>{value}<cr><lf>

Returns the serial baud rate.

97

Sdata

{7 or 8}

Sets the serial data bits.

98

Sdata?

<sp>{value}<cr><lf>

Returns the serial data bits number.

99

Sensor

C, X

Sets the active sensor where C = Primary or
Secondary and X is the turndown.

100 Sensor?

<sp>{C,X}<cr><lf>

Returns the active sensor as above.

101 Sensor.acc?

<sp>{value}<cr><lf>

Returns the accuracy of the sensor.

102 Sensorid?

<sp>{Address}<sp>MENSOR
,<sp>600SNSR,<sp>{Serial
No.},V{V.VV}<cr><lf>

Returns the active sensor’s serial number and
firmware version.

103 Setpt

{A value inside the upper/low-
er limits, and inside the range
of the active transducer and
turndown}

Sets the control setpoint for the active channel.

104 Setpt?

<sp>{value}<cr><lf>

Returns the control setpoint of the active chan-
nel in current units.

105 Span

{desired pressure or ?}

Sets span on active transducer (must be >50%
FS, and within 1% limit), or ? to clear previous
value. (See “Caldisable” command above ref#
19)

106 Span?

<sp>{value}<cr><lf>

Returns span scale factor for active transducer
and turndown.

107 Sparity

{Even, ODD , NONE}

Sets the serial parity.

108 Sparity?

{<sp>{string}<cr><lf>

Returns the serial parity.

109 Srqmask

Stable, Error or both

Sets the CPC 6000 to issue a service request
(SRQ) over the IEEE-488 when the pressure
control is stable, or an error occurs. These are
80 hex and 40 hex respectively.

110 Srqmask?

<sp>{string}<cr><lf>

Returns “stable”, “error” or “error, stable” de-
pending on the SRQ.

111 Sstop

{1 or 2}

Sets the serial stop bits.

112 Sstop?

<sp>{value}<cr><lf>

Returns the serial stop bits.

113 Stable?

<sp>{Yes or No}<cr><lf>

Returns YES if instrument is stable or NO.

114 Stabledelay

{0 to 65535}

Sets the stable time to the number of seconds
specified.

115 Stabledelay?

<sp>{value}<cr><lf>

Returns the stable time.

116 Stabletime

{0 to 65535}

Sets the stable time to the number of seconds
specified.

117 Stabletime?

<sp>{value}<cr><lf>

Returns the stable time.