WIKA CPC6000 User Manual
Page 55
![background image](/manuals/465881/55/background.png)
Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
55
75
Peakmax?
<sp>{value}<cr><lf>
Returns the maximum pressure since peakreset
was sent.
76
Peakmin?
<sp>{value}<cr><lf>
Returns the minimum pressure since peakreset
was sent.
77
Peakreset
none
Resets the peak values.
78
Port
{value}
Sets the Ethernet port of the instrument.
79
Port?
<sp>{value}<cr><lf>
Returns the Ethernet port of the instrument.
80
Precision
none
Sets the active channel to the precision mode:
.003% stablewindow and 4 second stabledelay.
81
Ptype
{Absolute or Gauge or A or G} Sets the instrument pressure type – non-native
type only works if the optional barometric sen-
sor is installed.
82
Ptype?
<sp>{Absolute
orGauge}<cr><lf>
Returns “Absolute” or “Gauge” for the pressure
type.
83
RangeMax?
<sp>{value}<cr><lf>
Returns the maximum range of the active trans-
ducer and turndown in the current units.
84
RangeMin?
<sp>{value}<cr><lf>
Returns the minimum range of the active trans-
ducer and turndown in the current units.
85
Rate?
<sp>{value}<cr><lf>
Returns the rate reading of the instrument in
current units/second.
86
Rdecpt?
<sp>{value}<cr><lf>
Returns the number of rate decimal points for
the active channel. See: Resolution
87
Reference
{EXTVAC or ATM}
Sets the reference type for vacuum pump
connected to the reference, or vented to atmo-
sphere. This option is only available on native
gauge units.
88
Reference?
<sp>{EXTVAC or
ATM}<cr><lf>
Returns the current reference type of the active
channel. See: Reference
89
Release?
{15 char string}
Returns: <sp>(YES or NO), {string}<cr><lf>
This command is used to release control of the
instrument when it is accessed by more than
one computer. To release control of the instru-
ment, {string} must be identifier used for the
“Acquire” command. Yes if release is successful.
No if instrument is being controlled with another
computer. Return {string} = identifier of the con-
trolling computer or AVAILABLE. See: Acquire?
and Unlock.
90
Repeat
None
Repeats output continually over serial port only.
91
Resolution
{4 to 6}
Sets the number of significant digits. See: decpt
92
Resolution?
<sp>{value}<cr><lf>
Returns the number of significant digits.
93
Rsetpt
{value}
Sets the rate setpoint of the active channel in
current units per second. Must be a value inside
the current sensor range.