Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

55

75

Peakmax?

<sp>{value}<cr><lf>

Returns the maximum pressure since peakreset
was sent.

76

Peakmin?

<sp>{value}<cr><lf>

Returns the minimum pressure since peakreset
was sent.

77

Peakreset

none

Resets the peak values.

78

Port

{value}

Sets the Ethernet port of the instrument.

79

Port?

<sp>{value}<cr><lf>

Returns the Ethernet port of the instrument.

80

Precision

none

Sets the active channel to the precision mode:
.003% stablewindow and 4 second stabledelay.

81

Ptype

{Absolute or Gauge or A or G} Sets the instrument pressure type – non-native

type only works if the optional barometric sen-
sor is installed.

82

Ptype?

<sp>{Absolute
orGauge}<cr><lf>

Returns “Absolute” or “Gauge” for the pressure
type.

83

RangeMax?

<sp>{value}<cr><lf>

Returns the maximum range of the active trans-
ducer and turndown in the current units.

84

RangeMin?

<sp>{value}<cr><lf>

Returns the minimum range of the active trans-
ducer and turndown in the current units.

85

Rate?

<sp>{value}<cr><lf>

Returns the rate reading of the instrument in
current units/second.

86

Rdecpt?

<sp>{value}<cr><lf>

Returns the number of rate decimal points for
the active channel. See: Resolution

87

Reference

{EXTVAC or ATM}

Sets the reference type for vacuum pump
connected to the reference, or vented to atmo-
sphere. This option is only available on native
gauge units.

88

Reference?

<sp>{EXTVAC or
ATM}<cr><lf>

Returns the current reference type of the active
channel. See: Reference

89

Release?

{15 char string}

Returns: <sp>(YES or NO), {string}<cr><lf>
This command is used to release control of the
instrument when it is accessed by more than
one computer. To release control of the instru-
ment, {string} must be identifier used for the
“Acquire” command. Yes if release is successful.
No if instrument is being controlled with another
computer. Return {string} = identifier of the con-
trolling computer or AVAILABLE. See: Acquire?
and Unlock.

90

Repeat

None

Repeats output continually over serial port only.

91

Resolution

{4 to 6}

Sets the number of significant digits. See: decpt

92

Resolution?

<sp>{value}<cr><lf>

Returns the number of significant digits.

93

Rsetpt

{value}

Sets the rate setpoint of the active channel in
current units per second. Must be a value inside
the current sensor range.