16
BR?
<sp>{value}<cr><lf>
Returns the B channel rate.
17
BRS?
<sp>{YES or NO}<cr><lf>
Returns the B channel rate stable flag.
18
BS?
<sp>{YES or NO}<cr><lf>
Returns the B channel stable flag.
19
Caldisable
{YES or NO}
Disables the zero and span commands (default
= YES). When the cal-disable is YES, the zero
and span commands are disabled.
20
Caldisable?
<sp>{YES or NO}<cr><lf>
Returns whether or not calibration of the active
sensor is disabled.
21
Cerr
none
Clears the error queue.
22
Chan
{A or B}
Sets the active channel on the instrument.
23
Chan?
<sp>{A or B}<cr><lf>
Returns which channel is active.
24
Chanfunc
<sp>{Press,peak,rate,baro}<
cr> <lf>
Sets the alternate function mode of the active
channel.
25
Chanfunc?
<sp>{string}<cr><lf>
Returns the alternate function mode of the
channel.
26
Control
none
Instrument placed in Control mode.
27
Control?
<sp>{YES or NO}<cr><lf>
Returns YES if active channel is in Control
mode; NO if otherwise.
28
Crate
{Slow, Medium, Fast, Vari-
able} for mid or high range
regulator, {Slow or Fast} for
low range regulator.
Sets the control slew rate for the current chan-
nel. For Variable rate see: Rsetpt command.
29
Crate?
<sp>{string}<cr><lf>
Returns the current rate - value corresponds to
the parameters for the CRATE command.
30
Decpt
<sp>{4, 5 or 6}<cr><lf>
Sets the number of significant digits displayed.
31
Decpt?
<sp>{value}<cr><lf>
Returns the number of significant digits dis-
played for the active channel.
32
Default
none
Sets the default values. This command takes
approximately 20 seconds.
33
DIO
{0 or 2}
0 turns the digital IO output pin off or low.
2 turns the digital IO output pin on or high.
34
DIO?
{value}
0 if input is low and output is low.
1 if input is high and output is low.
2 if input is low and output is high.
3 if input is high and output is high.
35
DOC
<mm/dd/yyyy>
Sets the date of cal for the active sensor and
turndown.
36
DOC?
<sp>{mm/dd/yy}<cr><lf>
Returns the date of cal for the active sensor and
turndown.
37
DOM?
<sp>{mm/dd/yyyy}<cr><lf>
Returns the date of manufacture.
38
Error?
<sp>{string}<cr><lf>