Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

51

7.4.5 Commands and Queries

Table 7.4.5 lists all of the current CPC 6000 specific commands and queries.

i

Notice

Channel specific commands are sent to only the active channel.
See ‘CHAN’ command.

Optional emulation modes are available in which the CPC 6000 can emulate remote functions of differ-
ent brands of pressure controllers. Please contact Mensor for more details.

Table 7.4.5 - CPC 6000 Commands and Queries

Ref Command

Data

Function / Response

1

?

See “Output Format” text

Returns data per the current output format.

2

Acquire?

{15 char string}
Returns: <sp>(YES or
NO),{string}<cr><lf>

This command is used when multiple comput-
ers would like to control the instrument. Yes if
acquisition is successful, No if instrument is be-
ing controlled with another computer. The string
is an identifier for the controlling computer. See:
Release? And Unlock.

3

Address

{1 to 30}

Sets the GPIB Address.

4

Address?

<sp>{value}<cr><lf>

Returns the GPIB Address within 1 to 30.

5

A?

<sp>{value}<cr><lf>

Returns the A channel pressure reading.

6

AR?

<sp>{value}<cr><lf>

Returns the A channel rate.

7

ARS?

<sp>{YES or NO}<cr><lf>

Returns the A channel rate stable flag.

8

AS?

<sp>{YES or NO}<cr><lf>

Returns the A channel stable flag.

9

Autorange

{ON or OFF}

Toggles autorange function to its opposite state,
either ON or OFF (enabled or disabled).

10

Autorange?

<sp>{ON or OFF}<cr><lf>

Returns whether the autorange function is en-
abled or disabled.

11

Autozero

none

Re-zero all ranges that can measure the vented
pressure. These adjustments are not password
protected and are not saved through power
cycles. This command takes approximately 60
seconds.

12

Autozero?

S,T,X,X

Returns autozero data where S represents
state (responses can be 0 = complete, 1 = local
autozero, or 2 = remote autozero), T represents
the estimated remaining time to complete in
seconds, and X is a (0) character since this data
location is not used at this time.

13

Autozero-
abort

none

Aborts autozero.

14

Baro?

<sp>{value}<cr><lf>

Returns reading from barometric sensor.

15

B?

<sp>{value}<cr><lf>

Returns the B channel pressure reading.