WIKA CPC6000 User Manual

Page 96

Advertising
background image

Automated Pressure Calibrator
CPC 6000

96

Mensor/WIKA Operating Instructions - CPC 6000

10.8.4 2-Point Calibration

A two-point calibration adjusts both the offset and the slope of the active turndown. The low point
must be within 20% of the minimum range of the active turndown and the high point must be within
20% of the maximum range of the active turndown. Insure that the head correction is adjusted prop-
erly, see

Section 10.9

. From the unlocked calibration data screen, press the [2 Point Calibration] key

as shown in Figure 10.8.4A. The ‘Low Point Cal’ window illustrated in Figure 10.8.4B will appear. This
window shows the maximum and minimum acceptable values and the current reading. The [?] key
pops up a help dialog with instructions on how to use this window. Enter the current known pressure
in the ‘Desired Reading’ data window and press the [OK] key.

Figure 10.8.4A - [2 Point Calibration] Key

Figure 10.8.4B - Low Point Cal Window

The ‘High Point Cal’ window illustrated in Figure 10.8.4C will appear. It’s function is identical to the
‘Low Point Cal’ window. Enter the new known pressure and press the [OK] key.

The [Cancel] key will exit either window and not make any adjustments to the calibration of the active
turndown.

The ‘Date of Calibration’ window illustrated in Figure 10.8.4D will appear. Enter the date of calibration
in a mm/dd/yyyy format exactly, including the ‘/’ separation marks. Press the [OK] key and the new
data of calibration will be saved to the turndown.

Figure 10.8.4C - High Point Cal Window

Figure 10.8.4D - Date of Calibration Window

Advertising