WIKA CPC6000 User Manual

Page 57

Advertising
background image

Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

57

118 StableWin

{%fs value}

Sets the stable window as a %FS.

119 StableWin?

<sp>{value}<cr><lf>

Returns the stable window in % of Span.

120 Standby

none

Instrument placed in Standby mode, all sole-
noids de-energized.

121 Standby?

<sp>{Yes or No}<cr><lf>

Returns YES if the active channel is in Standby,
NO if otherwise.

122 Step

<sp>{value inside upper and
lower limits and inside the
range of the active sensor
and turndown}<cr><lf>

Sets the control step size for the instrument.

123 Step-

{none or value}

Jogs the setpoint down one step. If a value is
entered, sets the step size to that value.

124 Step+

{none or value}

Jogs the setpoint up one step. If a value is en-
tered, sets the step size to that value.

125 Step?

<sp>{value}<cr><lf>

Returns the control step for the instrument.

126 Stop

none

Stop repeating output over serial port. See:
Repeat

127 Units

{units code, or output format
text from Measurement Units
Table 11.1 in the Appendix
Section}

Sets the instrument engineering units.

128 Units?

<sp>{string}<cr><lf>

Returns the instrument units in a text string.

129 Unlock

none

Releases Acquire locks. See: Acquire? and
Release?

130 UpperLimit

{Value inside primary trans-
ducer range}

Sets the upper control limit for the active chan-
nel.

131 UpperLimit?

<sp>{value}<cr><lf>

Returns the upper control limit for the active
channel.

132 Vent

none

Instrument placed in Vent mode.

133 Vent?

<sp>{Yes or No}<cr><lf>

Returns YES if active channel is in Vent mode,
No if otherwise.

134 Versions?

<sp>{string}<cr><lf>

Returns the versions of the firmware for the
instrument, gpib driver, gate array, installer, and
the graphics library.

135 Volume

{value in cc}

Sets the system volume in cc’s – only applicable
if the active channel is a low pressure regulator.

136 Volume?

<sp>{value}<cr><lf>

Returns the current system volume in cc’s – only
applicable if the active channel is a low pressure
regulator.

137 Zero

{desired pressure or ?}

Sets active sensor and turndown’s zero to en-
tered pressure, or ? to clear previous value. (See
“Caldisable” command above ref# 19)

138 Zero?

<sp>{value}<cr><lf>

Returns zero offset value for active sensor and
turndown.