Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
57
118 StableWin
{%fs value}
Sets the stable window as a %FS.
119 StableWin?
<sp>{value}<cr><lf>
Returns the stable window in % of Span.
120 Standby
none
Instrument placed in Standby mode, all sole-
noids de-energized.
121 Standby?
<sp>{Yes or No}<cr><lf>
Returns YES if the active channel is in Standby,
NO if otherwise.
122 Step
<sp>{value inside upper and
lower limits and inside the
range of the active sensor
and turndown}<cr><lf>
Sets the control step size for the instrument.
123 Step-
{none or value}
Jogs the setpoint down one step. If a value is
entered, sets the step size to that value.
124 Step+
{none or value}
Jogs the setpoint up one step. If a value is en-
tered, sets the step size to that value.
125 Step?
<sp>{value}<cr><lf>
Returns the control step for the instrument.
126 Stop
none
Stop repeating output over serial port. See:
Repeat
127 Units
{units code, or output format
text from Measurement Units
Table 11.1 in the Appendix
Section}
Sets the instrument engineering units.
128 Units?
<sp>{string}<cr><lf>
Returns the instrument units in a text string.
129 Unlock
none
Releases Acquire locks. See: Acquire? and
Release?
130 UpperLimit
{Value inside primary trans-
ducer range}
Sets the upper control limit for the active chan-
nel.
131 UpperLimit?
<sp>{value}<cr><lf>
Returns the upper control limit for the active
channel.
132 Vent
none
Instrument placed in Vent mode.
133 Vent?
<sp>{Yes or No}<cr><lf>
Returns YES if active channel is in Vent mode,
No if otherwise.
134 Versions?
<sp>{string}<cr><lf>
Returns the versions of the firmware for the
instrument, gpib driver, gate array, installer, and
the graphics library.
135 Volume
{value in cc}
Sets the system volume in cc’s – only applicable
if the active channel is a low pressure regulator.
136 Volume?
<sp>{value}<cr><lf>
Returns the current system volume in cc’s – only
applicable if the active channel is a low pressure
regulator.
137 Zero
{desired pressure or ?}
Sets active sensor and turndown’s zero to en-
tered pressure, or ? to clear previous value. (See
“Caldisable” command above ref# 19)
138 Zero?
<sp>{value}<cr><lf>
Returns zero offset value for active sensor and
turndown.