WIKA CPC6000 User Manual

Page 94

Advertising
background image

Automated Pressure Calibrator
CPC 6000

94

Mensor/WIKA Operating Instructions - CPC 6000

3.

Change the system password;

4.

Autozero all turndowns at one time.

Notice the padlock icon on the display. This signifies that the indicated values on the screen are not
able to be modified until the password has been entered.

[Unlock] Key: Touch [Unlock] (Figure 10.8.2A) to move on to the password screen. This function is
password protected. Details on this function are included in the

Calibration section

.

Figure 10.8.2A - Calibration Screen (locked)

Figure 10.8.2B - Calibrate Screen (unlocked)

[Autozero] Key: Press this key to have the CPC 6000 automatically re-zero all turndowns that can
measure the vented pressure (Figure 10.8.2C).

Figure 10.8.2C - [Autozero] Key

Figure 10.8.2D - Autozero Window

For an absolute CPC 6000, this automatic function will:

1.

Vent the system.

2.

Select the turndown with the smallest span that can measure the vented pressure, or read the inter-
nal barometric reference if installed.

3.

Record that pressure as “current barometric pressure”.

4.

Adjust the zero offset of each turndown so the turndown’s output equals the “current barometric
pressure”.