WIKA CPC6000 User Manual

Page 61

Advertising
background image

Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

61

18

_pcs4 func stby <unitno>

Instrument placed in standby mode in <unitno> units.

19

_pcs4 func vent <unitno>

Instrument placed in vent mode in <unitno> units.

20

_pcs4 id?

Returns instrument ID.

21

_pcs4 opt?

Returns instrument options.

22

_pcs4 outform <digit>

Sets output format.

23

_pcs4 outform?

Returns the current output format.

24

_pcs4 rangemin?

Returns the minimum pressure of the active transducer.

25

_pcs4 span?

Returns the stored multiplication factor from the active
transducer and turndown.

26

_pcs4 stabledelay <value>

1 to
255

Sets the number of consecutive readings that the pres-
sure must remain within the stable window for a pres-
sure stable indication.

27

_pcs4 stabledelay?

Returns the number of readings that must be within the
stable window before a stable pressure is indicated.

28

_pcs4 stablewindow
<value>

Sets the pressure window that is used to indicate pres-
sure is stable.

29

_pcs4 stablewindow?

Returns the pressure tolerance allowed for a stable pres-
sure indication as a % of span of the active transducer.

30

_pcs4 stat?

Returns mode and stable flag status “mode, stable CR
LF”.

31

_pcs4 unit <unitno>

Sets the instrument to specified engineering units.

32

_pcs4 unit?

Returns the current engineering units and type of trans-
ducer (A, G, D).

33

_pcs4 xducer?

Returns the number of the currently active transducer.

34

_pcs4 zero?

Returns the stored zero offset of the active transducer
and turndown in the current pressure units.

7.4.7.2 PCS 200 Emulated Commands and Queries

Table 7.4.7.2 – PCS 200 Emulated Commands and Queries

Command

Data

Function / Response

CX

Control Pressure at last control point and units

C$nnnnnnnX

Control Pressure at n in units $

D#X

CAL POINT CONTROL MODE NOT SUPPORTED

EX

Clear Error/Clear Service Request

E?X

Return error code and clear error

F$nnnnnnn1X

Re-initialize; $,n ignored

F$nnnnnnn2X

RETURN CAL DATA NOT SUPPORTED

F$nnnnnnn3X

Return Unit ID string; $,n ignored

F$nnnnnnn5X

RETURN QPS TEMPERATURE NOT SUPPORTED