WIKA CPC6000 User Manual

Page 62

Advertising
background image

Automated Pressure Calibrator
CPC 6000

62

Mensor/WIKA Operating Instructions - CPC 6000

F$nnnnnnn6X

RETURN NULL METER READING NOT SUPPORTED

F$nnnnnnn7X

RETURN VACUUM GAUGE READING NOT SUPPORTED

F$nnnnnnn8X

Return Clock Reading (Time); $,n ignored

F$nnnnnnn9X

Return Pressure Control Limits; $,n ignored

MX

Measure Pressure in current pressure units

M$X

Measure Pressure in units specified by $

M$nnnnnnnX

Measure Pressure in units specified by $; n ignored

M$nnnnnnnnsX

Measure Pressure in units specified by $; n,s ignored

Q#X

SEQ FUNCTIONS NOT SUPPORTED

RX

Not supported

R0X

Return to Standard Output Format

R1X

Re-initialize DO NOT USE - sets CPC 6000 to command prompt.

R2X

RETURN CAL DATA NOT SUPPORTED

R3X

Return unit ID string

R5X

RETURN QPS TEMPERATURE NOT SUPPORTED

R6X

RETURN NULL METER READING NOT SUPPORTED

R7X

RETURN VACUUM GAUGE READING NOT SUPPORTED

R8X

Return Clock Reading (Time)

R9X

Return Pressure Control Limits (Min and Max)

SX

Standby Mode

S$X

Standby Mode; $ ignored

S$nnnnnnnX

Standby Mode; $,n ignored

S$nnnnnnnsX

Standby Mode; $,n,s ignored

U$X

Change Units to units specified by $

VX

Vent Mode in current units

V$X

Vent Mode in units specified by $

V$nnnnnnnX

Vent Mode in units specified by $; n ignored

V$nnnnnnnsX

Vent Mode in units specified by $; n,s ignored

ZX

Automatic re-zero