Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

39

The [>] key begins execution of the program at the first program line. It is also used to re-start a pro-
gram at the current line after a wait/pause program line.

The next key shows a black square. Pressing this key stops execution of the program and resets the
program to the first program line.

The [II] key pauses program execution at the current program line. If the [>] key is subsequently
pressed, the program begins execution at the next program line.

6.5 [Local] Label

On the bottom right of the operation screen (Figure 6.3B) is a ‘Local’ label to indicate that an RS-232,
ethernet, or IEEE-488 host is not trying to control the instrument at this time. The text of this label will
change to “Serial”, the IP address of the controller, or “GPIB” when the CPC 6000 receives a remote
command.

6.6 [Setup] Key

Touch the [Setup] tab on the top left or top right corner of the Operation screen (Figure 6.3B) and a
new display appears with another set of tabs across the top as shown in Figure 6.6.1A. These tabs are
labeled [Channel], [Sensor], and [Controller]. The tab and screen that was last accessed is active by
default. The top tabs are used to set up conditions for the related feature/ function as explained in the
following text.

6.6.1 [Channel] Setup

Press the [Channel] tab to access the Channel Setup page as shown in Figure 6.6.1A. From there the
user can select the optional display function to be [Pressure], which will show only the current reading,
or have it show [Peak], [Rate], or [Barometer] readings.

Figure 6.6.1A - Channel Setup Screen