WIKA CPC6000 User Manual

Page 37

Advertising
background image

Automated Pressure Calibrator

CPC 6000

Mensor/WIKA Operating Instructions - CPC 6000

37

To change the name of the program, press the [Name] key. This displays a keyboard screen shown in
Figure 6.4.1B. Enter the name of the program and press [OK] to return to edit the program steps.

Figure 6.4.1B - Keyboard Screen

The Program Line Edit Screen (Figure 6.4.1C) sets the function of each program line. Each program
line performs the function selected from the ones displayed in the left-most column.

Figure 6.4.1C - Program Line Edit Screen

[Sensor] key: This key selects which sensor and turndown to use for the program.

[Mode] key: The [Mode] key selects the operation mode of the CPC 6000.

[Units] key: This key selects the pressure units.

[Setpoint] key: This key allows the control pressure to be set with the key just to its right. Because of
the flexibility of the CPC 6000, it should be noted that to make the CPC 6000 control at this setpoint
there needs to be another program line that puts the CPC 6000 into the control mode.

Advertising