39

Errorno?

<sp>{string}<cr><lf>

Returns CPC 6000 error code and text.

40

Filter

{Off, Low, Normal, High}

Sets the reading filter.

41

Filter?

<sp>{string}<cr><lf>

Returns the reading filter.

42

Gasdensity

{value}

Sets the head pressure gas density in lb/ft^3.

43

Gasdensity?

<sp>{value}<cr><lf>

Returns the head pressure gas density in lb/ft^3.

44

Gastemp

{value}

Sets the head pressure gas temperature in °F.

45

Gastemp?

<sp>{value}<cr><lf>

Returns the head pressure gas temperature.

46

Gateway

{string e.g. nnn.nnn.nnn.nnn}

Sets the Ethernet gateway address.

47

Gateway?

<sp>{string}<cr><lf>

Returns the Ethernet gateway address.

48

Height

{value}

Sets the head pressure height in inches.

49

Height?

<sp>{value}<cr><lf>

Returns the head pressure height.

50

Highspeed

{Yes or No}

Sets the active channel to high speed mode:
.006% stablewindow and 1 second stabledelay.

51

ID?

<sp>MENSOR,600,{ssssss},{
v.vv}<cr><lf>

Returns the instrument identity where {ssssss}
is the serial number, and {v.vv} is the software
version number.

52

IP

{string e.g. nnn.nnn.nnn.nnn}

Sets the IP address of the instrument.

53

IP?

<sp>{string}<cr><lf>

Returns the IP address of the instrument.

54

Keylock

{Yes or No}

YES to lock, or NO to unlock the on-screen
keys.

55

Keylock?

<sp>{Yes or No}<cr><lf>

Returns current keylock status as YES or NO.

56

List?

<sp>Pri,X,X;Sec,X,X;Bar,1<c
r><lf>

Returns list of available turndowns on installed
sensors in the active channel. X will be non-ex-
istent if the turndown is not available.

57

Listcal?

<sp>PRI,{sn},{td},{mm/
dd/yyyy},{td},{mm/dd/
yyyy};SEC,{td},{mm/dd/
yyyy},{td},{mm/dd/yyyy}<cr>
<lf>

Returns the serial number of each installed
transducer and calibration dates for each range.

58

Listrange?

<sp>PRI,{td},{min},{max},{td},
{min},{max};SEC,{td},{min},{m
ax},{td},{min},{max}<cr><lf>

Returns the minimum and maximum ranges of
all installed sensors for the active channel.