Environment settings parameters, Output method, Output method environment settings parameters – KEYENCE LT-9001 Series User Manual

Page 147

Advertising
background image

8-17

Outputting Measured Values by Using External
Synchronization

This section describes how to synchronize the measured value with the measurement performed by the

commands from the input terminal (TIMING input) and output it from the RS-232C interface.

This method does not use any incoming commands so it is used when the measured value is only

imported or when the unit is connected to the device that cannot send the command, such as a data

logger.

Outputting the measured value and changing the settings by using the command are possible even when the

measured value output with external synchronization is in use.

Avoid triggering the next measured value while the current measured value is being output via RS-232C. That

next measured value will not be output under these conditions.

Sets the environment settings as follows.

Refer to “RS-232C Interface (page 6-5)” for the environment settings.

Outputs the measured value fixed by the external synchronization (TIMING input).

Output contents differ depending on the D-MODE setting as shown below.

Output Method

Environment Settings Parameters

Note

Reference

E LT-9001-IM

8

Chapter

RS-232C

Item

Parameter Remarks

BAUDRATE

9600/14400/19200/38400/
57600/115200 bps

PARITY

NONE/EVEN/ODD

STOPBIT

1/2

D-LENGTH

7/8

D-MODE

S0/S1/S2

Select one of three options shown in the left
column.

Set the parameter according to the external
devices to be connected.

D-MODE

Output contents

OFF

Does not output the measured values in
external synchronization.

S0

Outputs OUT1 and OUT2 at the same time.

S1

Outputs only OUT1.

S2

Outputs only OUT2.