Basic settings, Area – KEYENCE LT-9001 Series User Manual

Page 79

Advertising
background image

Basic Settings

The procedure used to set the ranges of the areas is

explained as follows.

1

Select [AREA] from [BASIC] on the program

mode screen.

The area setting screen appears.

2

Select [AREA 1].

3

Select a measurement method for [AREA1].

Select from among [AVG], [MAX], [MIN] and

[DIFF].

4

Select [AREA2].

Select a measurement method for [AREA2] in the

same manner.

5

Press the [ENTER] button.

The areas are fixed.

[DIME] (Area measurement) can only be set for Area2.

However, it cannot be set when a calculation (A1–A2

and A1+A2) is set for [MEASUREMENT].

Area setting for the area measurement

Switch the upper and lower limits of Area 1 and set

both ends of the scanning range in Area 1. Measure

the area of Area 2 using the approximated line of

Area 1 as a reference.

Reference

Note

4-13

This section describes area, scanning, dark, and

smoothing.

The following measurements can be performed

using the data that is obtained in the area

specified within the scanning range.

Two areas can be set: [A1 (Area 1)] and [A2 (Area

2)].

Using the respective values measured in [A1] and

[A2], following measurements can be calculated.

AVG

Measures an average value

DIFF

Measures a differential between the maximum
and minimum value

MIN

Measures a minimum value

MAX

Measures a maximum value

DIME

Measures an area (µm

2

or mil

2

)

AREA

E LT-9001-IM

4

Chapter

Operations

and

Fun

ction

Settings

in
the

Profile

Mode

Area setting range

DIME (Area measurement)

MAX (maximum value)

AVG (average value)

MIN (minimum value)

DIFF (differential value)