Appendices

E LT-9001-IM

A-3

Symptom

Check point

Remedy

Is the unit properly wired?

Check the output circuit and wiring and connect
properly.

Judgment cannot be output

Is the tolerance properly set up?

Perform the tolerance setting again.

properly.

(

Page 5-7)

Is the I/O mode properly set up?

Select comparator mode for I/O mode.

(

Page 5-20)

Is the unit properly wired?

Check the wiring and connect properly.

Does the voltage exceed

Set the analog scaling appropriately according to

+10.5 V or below -10.5V?

the measured value. (

Page 5-15)

Analog output cannot be output

Is an oscilloscope or a hi-speed A/D board

The resolution may decrease if these devices are

properly.

used?

used.
Change the analog scaling so that the effect
should be minimized.

Is the I/O mode properly set up?

Select binary mode for I/O mode.

(

Page 5-20)

Is the unit properly wired?

Check the output circuit and wiring and connect

Binary output cannot be output

properly.

properly.

Is the logic properly set up?

Negative logic is used in output.

Is the import timing properly set?

Use strobe output to enable importing in proper

timing. (

Page 7-4)

Is BIN SEL input properly input?

Input correctly. (

Page 7-4)

Is the communication cable properly

Connect the communication cable properly.

connected?

(

Page 8-2)

Does the type of the communication cable

Check the connection of the communication

conform to the connecting specifications?

cable. (

Page 8-2)

Cannot communicate via the

Are the communication specifications properly Set the communication specifications of the LT-

RS-232C.

set up?

9000 the same as that of the external devices.

(

Page 8-2)

Is the appropriate communication program

Check whether the commands and the delimiters

used?

are properly received or sent. (

Page 8-3)

Is the setting properly set up according to the

Set the D-MODE appropriately.

communication method or external devices?

(

Page 8-3)

The image of the measurement

Is the display mode appropriately set up?

Set it up appropriately.

target is not displayed.

(

Page 2-12)

Is the remote control console properly

Connect the remote control console properly.

The remote control console does connected?

(

Page 1-8)

not function.

Is Keylock set to ON?

Set the Keylock to OFF.

(

Page 5-5)