OUTPUT Settings

Auto peak hold [A.PEAK]

Measures continuously, and displays and outputs the measured value as needed.

• Auto bottom hold [A.BOTTOM]..........Displays and outputs the minimum value as needed that is measured

continuously.

• Auto peak-to-peak hold [A.P-P] .........Displays and outputs the differential between the maximum and

minimum value as needed that are measured continuously.

• Judgment outputs are output according to the measured value.

• The ON status of each output means that the NPN open-collector output is ON.

• The ON status of each input means that the input and the COM terminal are short-circuited.

• The RS-232C output is synchronized with the timing input when the "D-MODE" in the environment setting is set

to any option other than OFF. (

page 8-17)

Refer to "Binary Output" (page 7-10) for details.

• If setting the RESET input to ON when the TIMING input is ON, the judgment standby status is displayed ("- - -

-" is indicated) until the TIMING input becomes OFF.

• The TIMING input and RESET input can be controlled using the RS-232C. (

Page 8-6)

• The TIMING input and the RESET input can be controlled by using the [HOLD] button and the [ESCAPE]

button respectively. (

Page 5-3)

Reference

Judgment

Measured value

Judgment standby status