YSI 6000UPG User Manual

Page 103

Advertising
background image

6-3

deviation may also be due to fouling of the platinum ORP sensor. This problem can usually be
corrected by cleaning the platinum ring according to the manufacturer’s instructions.

To determine whether the sensor is functioning correctly, place the ORP probe in 3682 Zobell
solution and monitor the millivolt reading. If the probe is functioning within specifications, the
ORP reading should be within the range of 210-250 mv at normal ambient temperature. If the
reading is outside of this range, the probe should be cleaned, rechecked, and then can be calibrated
to the correct value (231 mV at 25

o

C) using the calibration procedure outlined in Section 4.2.

ORP readings for the same solution can vary up to 100 mv depending on the temperature.
However, no standard compensation algorithms exist for this parameter. Be sure to take this factor
into account when reporting ORP values and checking sensor calibration. For Zobell solution,
consult the following chart:

TEMPERATURE, CELSIUS

ZOBELL SOLUTION VALUE, MV

-5

270

0

263.5

5

257

10

250.5

15

244

20

237.5

25

231

30

224.5

35

218

40

211.5

45

205

50

198.5

MEASUREMENT AND CALIBRATION PRECAUTIONS

(1) Instructions for preparation of ORP calibrator solutions can be found in Section 2580 A. of
Standard Methods for the Examination of Water and Wastewater. Alternatively, the reagent is
available from YSI.

(2) ORP readings usually stabilize much more rapidly in standards (e.g., Zobell solution) than in
most environmental water. Be certain to consider this factor when determining ORP in field
studies.

(3) Clean and store the pH/ORP sensor following the instructions in Section 7 and Appendix G of
this manual.