8-2

COMMUNICATION PROBLEMS

SYMPTOM

POSSIBLE CAUSE

ACTION

REF

Can not communicate
with sonde

Comm port not properly selected

Change comm port

5.5

Cable connection is loose

Check both ends of cable; secure
connectors

Sonde batteries are depleted

Change batteries

7.1

Water in battery compartment

Dry compartment

7.1

Internal failure

Return sonde for service

C

Scrambled data

Unmatched baud rate between
host and sonde

Match the baud rate

2.4
4.5

Host is too slow

Use faster computer or different
software

Interface cable failure

Check cable for damage. If
necessary, return for service

Internal failure

Return sonde for service

C

Unable to upload data
from sonde to PC

Insufficient PC disk space

Make space or specify different
drive

5.7

Communication problem

See communication sections above

Internal failure

Return sonde for service

C

SENSOR PERFORMANCE PROBLEMS

SYMPTOMS

POSSIBLE CAUSE

ACTION

REF

Oxygen unstable or
inaccurate

Probe not properly calibrated

Follow DO cal procedures

3.1

Membrane not properly installed
or may be punctured

Follow 6030 setup procedure

2.2

DO probe electrodes require
cleaning

Follow DO cleaning procedure

7.2

Water in probe connector

Dry connector; reinstall probe

2.2

Algae or other contaminant
clinging to DO probe

Rinse DO probe with clean water

Barometric pressure entry is
incorrect

Repeat DO cal procedure

3.1

Cal at extreme temperature

Recal at (or near) sample
temperature

DO Charge too high

Follow DO cleaning procedure

7.2

DO probe has been damaged

Replace 6030 probe

2.2

Internal failure

Return sonde for service

C